Uploading pump data erases data uploaded from other pump
In a two pump configuration, where a suspended 640G is serving as cgm data collector only, the uploaded syncs the suspended state to Nightscout, invalidating
- site changes
- basal rate
- bolus
- carb data
The alternating upload from loop and uploaded competes for the data, resulting in alternating displays of metadata and cgm-only, and data loss on the loop frontend to boot (which sometimes fails to handle such a situation where the data has been deleted on nightscout only - also an issue with using the careportal menu on the Nightscout site)
- 1730 a Dexcom G5 sensor got started in addition to the running Enlite
- 1930 Dexcom got calibrated
- slight deviations in the tracks of both sensors leading to "loops". That's not a bug, the data source is linked to the data.